目的:观察免疫放射分析与微粒子化学发光法检验血清促甲状腺素(TSH)的价值。方法:分别采用免疫放射分析与微粒子化学发光法检测40份血清标本的TSH水平。结果:免疫放射法、微粒子化学分析法测量的TSH水平分别为(7.889±0.355)、(8.077±0.299)μIU/mL,两者比较,差异无统计学意义(t=1.532,P>0.05)。两种方法测定TSH值具有较好的相关性(γ=0.99)。结论:免疫放射分析与微粒子化学发光法检验血清TSH水平具有较好相关性。
Abstract:Objective:To determine the diagnostic value of immunoradiometric and chemiluminescence assay for assessment of serum thyrosin stimulating hormone(TSH).Methods:Forty serum samples were subject to measurement of TSH by using the two different techniques.Results:Immunoradiometric and chemiluminescence assay yielded an unmarked difference in the detected level of TSH[(7.889 ± 0.355) vs(8.077 ±0.299) μIu/mL,P >0.05].The measured TSH values correlated well between the two techniques(r= 0.99) Conclusion;The measured TSH values correlated well between the two techniqugs.
[1]杨国昌,许叶,张抚,等.精密度评价和方法比较中NCCLS评价方案的应用[J].临床检验杂志,1997,17(1):47.
[2]Spencer CA,Takeuchi M.Current status and performance goals for serum thyrotropin(TSH)assay[J].Clin chem,1996,42(1):140-145.
[3]Gosling JPA.Decade of development in immunoassodology[J].Clin Chem,1990,56(8):1408-1427.
[4]OgnibeneA,Drark CJ,Jeng KY,et al.A new modular chemiluminescence immunoassay analyser evaluated[J].Clin Chem lab Med,2000,38(3):251-260.
基本信息:
中图分类号:R446.6
引用信息:
[1]冯玉俊.免疫放射分析与微粒子化学发光法测定血清TSH的价值[J].广州医学院学报,2013,41(05):67-68.