摘要:

目的:对日本东亚公司的SYSMESE-9000全自动五分类血液细胞分析仪进行评价并报告。方法:用手工方法和库尔特JT-IR血液细胞分析仪测定与SYSMEX SE-9000血液细胞分析仪测定进行比较分析。结果:批内、日间精密度:(CV)WBC<2.4%,RBC<0.7%,HGB<0.7%,PLT<2.2%,其他各参数均<1%;线性试验:WBC在(2.35~179.87)× 109/L,RBC在(0.26~12.68)× 109/L,PLT在(10-962 × 109/L)之间,有较好的线性;携带污染率<0.1%;血液参数对比实验,各CBC参数与COULTERJT-IR进行对照、结果有较好的相关性;白细胞分类评价,中性粒细胞,淋巴细胞,单核细胞,嗜酸粒细胞,嗜碱粒细胞相关系数(r)分别为 0.9666,0.9556,0.7509,0.7438,0.6161。结论:SYSMEX SE-9000全自动五分类血液细胞分析仪是一种性能较为优良,值得推广应用的血液细胞分析仪。

Abstract:

ve: To evaluate the instrument of SYSMEX SE -9000 Automatic Hematology analyzer. Methods: Blood samples were analyzed with manual method, coulter JT - IR hematology analyzer and SYSMEX SE - 9000 hematology analyzer and the results were compared. Results: The precisions of in-run and between-run were: WBC< 2. 4% , RBC < 0. 7% , HGB < 0. 7% , PLT < 2. 2% , the others < 1 % ; Linearity test showed that within the ranges as follow: WBC(2.35 -179.87) ×109/L, RBC (0.26-12.68) ×1012/L, HGB (6-375g/L), PLT(10-962)× 10 /L,the linearity was good. The rate of cross interference of sample was < 0. 1%. In the comparison of menotology parameters, the results showed that there was a good correlation between the results from SYSMEX SE -9000 and coulter JT - IR analyzer. In the evaluation of leukocyte classification, the r was 0. 9556,0. 9666, 0.7509, 0.6161,0.7438 respectively for granulocytes, lymphocytes mononuelears, eosinophils and basophils. Conclusion; SYSMEX SE -9000 is a good automatic hematology analyzer and is worth of recommendation.

参考文献

[1]中华人民共和国卫生部医政司主编.全国临床检验操作规程[M].第2版.北京:人民卫生出版社,1994.5

[2] 丛玉隆主编.今日临床检验学[M].第1版.北京:中国科学技术出版社,1998,12-13.

[3] 孙朝晖,朱爱辛,林龙顺,等.MET BC-200血液细胞分析仪评价试验报告[J] .实用医学杂志,2000,16(8);698—701.

[4] 孙朝晖,潘芳,林龙颇,等.DANNA EXCELL-18血液细胞分析仪评价试验报告[J] .现代医学仪器与应用,2001,13(1):19—21.

[5] 潘柏申,王玉亭,林镇海,SYSMEX SF-3000血液细胞分析仪应用评价[J].中华医学检验杂志,1998,21(5):318~319.

基本信息:

中图分类号:R446.11

引用信息:

[1]孙朝晖,潘芳,姜小琴,张速林,杨永泉,林龙顺,苏健.东亚SYSMEX SE—9000全自动五分类血液细胞分析仪评价试验报告[J].广州医学院学报,2001(04):44-47.

检 索 高级检索

引用

GB/T 7714-2015 格式引文
MLA格式引文
APA格式引文