摘要:

目的:验证和评价雅培i2000SR全自动化学发光仪定量检测鳞状细胞癌抗原(SCC Ag)的分析性能。方法:对雅培i2000定量检测SCC Ag的精密度、线性进行验证。用i2000和酶联免疫吸附技术(ELISA)方法检测相同标本SCC Ag,对检测结果进行比较分析。结果:精密度均符合雅培性能标准;SCCAg检测呈一次线性(r=0.999,P<0.05),线性范围为:0.24-61.9 ng/mL;两种方法检测相同标本,i2000的阳性率为60%,ELISA的阳性率为35%。以i2000为标准,ELISA的相对敏感度为54.2%、相对特异性为93.8%,两种方法的总符合率为70%。结论:雅培i2000检测SCC Ag的主要分析性能验证结果与厂商声明的基本一致,与旧方法ELISA的总符合率良好,符合质量目标要求,可应用于临床标本检测。

Abstract:

Objective:To evaluate the analytical performance of i2000SR automatic chemiluminescence analyzer in quantitative detection of squamous cell carcinoma antigen(SCC-Ag).Methods:We validated the precision and linearity performance of i2000SR(Abbott) analyzer in quantitative measurement of SCC-Ag.The same specimens were tested by both i2000 and ELISA,and then the results were compared and analyzed. Results:The precisions were in line with the Abbott performance standards.SCC-Ag showed good linearity(r = 0.999,P <0.05 ) with a range between 0.24 and 61.9 ng/mL.In the detection of same specimens,the positive rate for i2000 was 60%vs 35%for ELISA.Using i2000 test as the reference,the relative sensitivity and specificity of ELISA was 54.2%and 93.8%,respectively.The total coincidence rate of the two methods was 70%.Conclusion:Performance validation of i2000SR analyzer in detection of SCC-Ag basically correlates with as claimed by the manufacturer and is well consistent with conventionally used ELISA.Therefore,i2000SR meets quality standards and can be applied for clinical purposes.

参考文献

[1]张葵.定量检测系统方法学性能验证实验的基本方法[J].临床检验杂志,2009,27(5):321-323.

[2]Clinical and Laboratory Standards Institute.EP5-A2. Approved Guideline-Second edition.2004.Evaluation of precision performance of quantitative measurement methods [S].USA:CLSI,2004.

[3]Clinical and Laboratory Standards Institute.EP6-A2. Approved Guideline-Second edition.2003.Evaluation of the linearity of quantitative measurement procedures[S]. USA:CLSI,2003.

[4]戎奇吉,戴金华,马建波,等.101例肺癌患者血清肿瘤相关抗原Scc-Ag检测结果分析[J].中国卫生检验杂志,2008,18(7):1363-1364.

[5]张丽民.化学发光标记及发光免疫分析[J].基础医学与临床,1995,15(4):247-250.

[6]徐建华,黄宪章,庄俊华,等.罗氏Modular全自动生化分析仪酶学指标检测性能验证[J].检验医学,2010,25 (2):81-85.